<16. COMMISSIONING OF SENCOM pH/ORP>

16-17

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

pH

Time

Good Electrode

Bad Electrode

tR : Recovery Time

tW : Wash Time

tI : Wash Period

1/2∆pH

∆pH

1/3tR

tI : Wash Period

tW : Wash Time

tR : Recovery Time

F050605_3.ai

Figure 16.5

16.6.6

Factory setup

For “Factory setup,” there is no item to be set by the user.

NOTE

This menu is for service engineers only. This section is protected by a password. Attempting to

change data in the factory adjustment menu without the proper instructions and equipment could

result in corruption of the instrument setup and damage the performance of the unit.

16.7 Display setup

This screen is used to make various settings relating to screen display.

NOTE

Settable items differ depending on settings in “Sensor setup” and “Measurement setup.”

16.7.1

Main display

Three measurement values can be set to display on the Main display as a primary value (1st

line), a second value (2nd line) and a third value (3rd line) respectively.
On the “Additional text”, a text of up to 12 alphanumeric characters can be assigned to each

measurement value.
Additional texts are displayed on the Main display, and are useful for identifying measurements.

In some cases, not all 12 characters can be displayed due to the letters; check the texts

displayed on the Main display after setting. If a part of the text is missing, adjust the number of

characters.