<Appendix 3 For ISC (Inductive Conductivity)>

App.3-11

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

The conductance (1/R) is proportional to the specific conductivity and a constant factor that

is determined by the geometry of the sensor (length divided by surface area of the hole in the

toroid) and the installation of the sensor.
There are 2 toroids mounted in the “dough nut” shaped sensor. The liquid also flows through the

second toroid and therefore the liquid turn can be considered as a primary wind ing of the second

ring transformer.
The current in the liquid will create a magnetic field in the second toroid. The induced voltage

being the result of this magnetic field can be measured as an output.
The output voltage of this “receive coil” is therefore proportional to the specific con ductivity of the

process liquid.

n

Changing the settings

If any setting is accidentally changed, values to the right of the relevant arrow in Table 6 are all

initialized.

Table 6 Parameters that initialize other values

Measurement ->

Output: Process

parameter ->

Linear: 0% value, 100% value
Table
Communication: HART: PV

Display setup: Main display
Trend Graph Screen ->

Y-axis (low, high)

Communication: HART

Configure sensor:

Measuuring unit ->

Display setup: Main display: unit

App.

ISC

Advertising