Nzero current, Nslope, Nsensor current – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 147: Navailable koh, Ncompensation, Nsensor wellness, Nlast calibrated, Ncalibration due, Do 12

Advertising
background image

<12. OPERATION OF DO (Dissolved Oxygen)>

12-7

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

n

Zero Current

= The offset value of a calibrated sensor. This is the offset of the sensor and sensor circuit in the

zero oxygen condition.

n

Slope

This parameter indicates the sensitivity of the sensor after calibration. It is indicated as a

percentage of the reference sensitivity based on the selection of the sensor or that has been

entered.

n

Sensor Current

= The raw output of the sensor before it is calibrated and temperature compensated.

n

Available KOH

This parameter indicates the amount of available KOH remaining in the sensor. This amount

is calculated and displayed by totalizing the amount of KOH consumed by dissolved oxygen

measurements.
When replacing the solution in the sensor, calibration should be made to reset (clear) diagnostic

sensor data.

n

Compensation

For compensation of measured value, salinity compensation and pressure compensation are

available. To use pressure compensation, enter the pressure level directly.

n

Sensor wellness

At the Sensor wellness window, the soundness of a module is displayed. A larger number of n

in each gauge indicates that the parameter concerned is sound. A gauge is indicated for only

those parameters whose sensor wellness setting is “enabled,” while a bar (----) is displayed if the

sensor wellness setting is “disabled.”
Sensor wellness setup can be made in Commissioning → Measurement setup → Sensor diag.

settings. For details, see section 13.2.7.
The “Reset wellness data” button can reset wellness data.
When a sensor or a membrane is exchanged or replaced, sensor wellness data should be reset.

NOTE

When a sensor is replaced, the replacement can be recorded manually into a logbook. (Refer to

the figure 12.7.)

n

Last calibrated

= date on which the last sensor calibration was performed. The displayed value of the Zero is the

result of this calibration. The displayed value of Slope was calibrated on this date only if the last

calibration was a 2-point calibration.

n

Calibration due

= the date when the calibration must be done next according to the settings of the calibration

interval. The calibration intervals are set in Commissioning → Measurement setup → Calibration

settings → Limits and timing.

DO

12

Advertising
This manual is related to the following products: