<10. COMMISSIONING OF ISC (Inductive Conductivity)>

10-5

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

NOTE

Perform a zero check for air calibration while the temperature compensation is set to NaCl.

l

c.c. high

High limit of the cell constant expressed as a % of the nominal value. During calibration this value

is used to check if the calibrated cell constant remains within reasonable limits.

l

c.c. low

Low limit of the cell constant expressed as a % of the nominal value. During calibration this value

is used to check if the calibrated cell constant remains within reasonable limits.

l

Step Range

Set the range over which the stability of a measured value is checked. If variations of a measured

value over the stabilization time are within this set value, the measured value is judged to have

stabilized.

l

Stabilization time

During calibration, the stability of the measurement is constantly monitored. When variations of

the value are within a value set in Step Range over this stabilization time set here, the value is

regarded as being stable. If the value does not stabilize within 10 minutes, calibration is aborted.

l

Calibration interval

Set the interval in which a new calibration must take place. If the interval set here is exceeded,

the instrument will be notified according to the setting in “Calib. time exceeded” in the error

configuration.

10.1.6 Concentration

The “Concentration” can be set only when “Conduct. + Concentr.” or “Concentration” is selected

for the Measurement in the Measurement setup menu.
The corresponding concentration given by a measured conductivity can be shown on the Main

display.

l

Concentration table

Concentration can be calculated by using the temperature compensation matrix or the additional

concentration table.

• By using the temperature compensation matrix

Set the “Additional table” to “Disabled”. The concentration can be obtained from the

temperature compensation matrix (based on the relation between the conductivity at the

reference temperature and the concentration).

Select Temp. compensation → Method, and then select Matrix. Next, choose a solution

or User defined 1 or 2 (see Section 10.1.4). The concentration cannot be obtained by any

other temperature compensation method (TC or NaCl).

• By using the additional concentration table

The additional concentration table indicates the relation between the conductivity and the

concentration at the reference temperature. Set the “Additional table” to “Enabled”. The

concentration can be obtained from the additional concentration table.

After entering values, select “Yes” for “Check values?”. Values are interpolated into empty

cells. If any mandatory cells are left blank, an error will be issued. An error message is also

issued if there is any error in the concentration table.

After completing entering values in the additional concentration table, set “Additional table”

to “Enabled”.