Supplement

Model FLXA21

2-Wire Analyzer

IM 12A01A02-01E 1/3

5th Edition

Thank you for selecting our FLXA21 2-Wire Analyzer.
Though User’s Manual, IM 12A01A02-01E 5th Edition, is provided with the product, an addition to

the manual has been made.
Please attach a copy of this supplement to the manual and follow the instructions below when

using the instrument.

Note

n

Instructions related to NEPSI type is added.

Model FLXA21 2-Wire analyzer has been certified as explosion proof equipment by NEPSI.
Information and instructions related to NEPSI certification is added to the manual.

l

Section 1.4 Regulatory Compliance

The followings are added to the article.

Explosion-proof (Intrinsically safe type and non-incendive) (for suffix code: -EA):

NEPSI Intrinsically safe approval

Applicable standard

GB 3836.1-2010 Explosive atmospheres-Part 1: Equipment - General requirements

GB 3836.4-2010 Explosive atmospheres-Part 4: Equipment protection by intrinsic safety “i”

GB 3836.20-2010 Explosive atmospheres-Part 20: Equipment with equipment protection level (EPL)

Ga

Type of protection

Ex ia IIC Ga

T4: for ambient temperature: -20°C to 55°C

T6: for ambient temperature: -20°C to 40°C

Atmosphere pressure: 80kPa (0.8bar) to 110kPa (1.1bar)

Degree of Protection of the Enclosure: IP66

Control Drawing (NEPSI types)

Measuring module 1

Note: The measuring module on this drawing means

the sensor module on this General Specifications.

Measuring module 2

Sensor 1

Sensor 2

(*)

(Note 1)

Safety Barrier

Housing Assembly

Supply -

Supply +

Non Hazardous Location

Hazardous Location

-

+

Electrical data are as follows;

Maximum Voltage (Ui) = 30V

Maximum Current (Ii) = 100mA

Maximum Power (Pi) = 0.75W

Internal Capacitance (Ci) = 13F

Internal Inductance (Li) = 0mH

Note 1: The output current must be limited by a resistor “R” such that Imaxout=Uz/R (linear source).

Note 2: Safety barrier certified by NEPSI should be used.

Note 3: When using non isolation barrier, connect (*) to IS earthing system.

Note 4: Sensor module 2 is installed when required.

When measuring inductive conductivity, only one module can be installed.