<11. CALIBRATION OF ISC (Inductive Conductivity)>

11-1

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

11. CALIBRATION OF ISC (Inductive

Conductivity)

Inductive conductivity meters must be calibrated after being installed in a measurement location

or moved to a different location.
Inductive conductivity meters do not generally need to be calibrated repetitively if they have been

calibrated once before the start of use. If the cell is severely contaminated or has been subject to

abrasion (possibly during cleaning), calibration may be necessary.

Execute air calibration first, if needed.
When using solutions, manual, automatic, and sample calibrations are available.

Execute:
Calibration
HOLD

Setup:
Commissioning
Change language
Start Quick Setup

Execute & Setup

Calibration

Cell constant(manual)
Cell constant(automatic)
Air calibration
Sample
Temperature coefficient
Temperature calibration

Figure 11.1

Calibration

NOTE

A default is “10 min” for “Auto Return”. When maintenance like a calibration that may take much

time is performed, “60 min” or “Disable” is recommended to be selected. (Refer to the section

10.6.3.)

Calibration is carried out by measuring a solution which has a known conductivity and adjusting

the instrument to show the correct conductivity value.

● “Cell constant (manual)”, “Cell constant (automatic)

A calibration solution can be prepared in the laboratory. A salt solution is prepared with a

known precise concentration. The temperature is stabilized to the reference temperature of the

instrument (usually 25°C). The actual conductivity of the solution is taken from literature data. The

typical calibration solutions include NaCl and KCl; see Appendix 3. To calibrate the instrument,

the sensor is removed from a process and suspended in the solution, the conductivity value from

the literature data is then entered and the calibration routine completed.
Make sure the sensor does not touch the sides of the container (see Figure 11.2).

● “Sample calibration”

This calibration uses a process solution. In this method, the conductivity of the process solution

under measurement is measured with a standard instrument. In this case, care must be taken to

make the measurement at the reference temperature since differences in the type of temperature

compensation of the instruments may cause an error. With this method, it is not necessary to

remove the sensor from the process. Because the sensor is immersed in the process, errors

caused by the installation characteristics are compensated for.

ISC

11