Wiring and installation, Wiring and installation -1, Ph sc isc do – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 28: Sencom

Advertising
background image

<2. WIRING AND INSTALLATION>

2-1

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

2. WIRING AND INSTALLATION

Install the cable glands into the FLXA21 before installing it (see section 2.3.) Upon delivery, the

cable glands are not installed.

(Plastic housing)

(Stainless steel

housing)

*2

*2

*4

*5

*6

*6

n n n n n n n

n n

n n n n n n n n n

+

*1: Use a 2-wire shielded cable with an outside diameter of 6 to 12 mm.

*2: Connect the analyzer to gland. (Class D ground: 100 ohm or less)

The way of connecting the grounding cable varies depending on the plastic housing and stainless steel

housing.

In the case of the plastic housing, connect the grounding cable to the terminal of the power module inside,

and in the case of the stainless steel housing, connect the grounding cable to the terminal of the housing.

Use a cable with an outside diameter of 3.4 to 7 mm for the grounding line of the plastic housing.

The minimum cross sectional area of the protective grounding cable should be 0.75 mm

2

.

*3: This line is connected to a distributor or 24V DC power supply.

*4: Terminal numbers for each sensor module are shown below.

*5: Two modules of the same kind of measurement/sensor type can be installed. When measuring inductive

conductivity or pH/ORP with the SENCOM sensor, only one module can be installed.

*6: The terminal box may be necessary depending on the sensor cable length and the distance between the

analyzer and the sensor.

The SENCOM sensor is to be connected directly to the analyzer without a terminal box.

*7: Two outputs, output 1 and output2, of PH201G or SDBT are same signals.

*1

WTB10 or BA10

Terminal box

FLXA21 2-Wire Analyzer

Sensor

WTB10 or BA10

Terminal box

Sensor

PH Module

SC Module

ISC Module

DO Module

SENCOM Module

16

15

19

17

13

18

14

12

11

PH

NC

18

17

14

13

15

16

12

11

DO

15

16

14

17

13

12

11

ISC

NC

16

15

14

13

12

11

SC

87

86

NC

84

83

82

NC

SENCOM

Power supply

Output 1 (1-5V DC)

20 to 130V DC

or

80 to 138V AC, 47 to 63Hz

Power supply

20 to 130V DC

or

80 to 138V AC, 47 to 63Hz

HOLD

WASH

FAIL

Output 2 (1-5V DC)

Output 1 (1-5V DC)

Output 2 (1-5V DC)

Case of Distributor

PH201G (Style B)

Case of Distributor

SDBT

(*3)

*3

*7

*7

(*3)

A

+

– CMN CMN

B

C

D

F

H

L N

A

+

+

B

1

+

2

F

H

L N

b

a

d

c

f

e

Figure 2.1

Wiring diagrams

PH

SC

ISC

DO

SENCOM

2

Advertising
This manual is related to the following products: