<15. OPERATION OF SENCOM pH/ORP>

15-2

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

15.1 Change language

The screen is set to display English at factory shipment; if you wish to use the FLXA21 in another

language, first select a language as described in section 2.7.

15.2 Quick setup

The Quick setup screen is used to set up the basic items you want to set up first, such as the

date/time and sensor settings. The detailed settings are described in chapter 16, Commissioning.
You may leave the Quick setup now and return to it later; however, it is recommended to perform

the quick setup first.
Each time the FLXA21 is started up, this screen is displayed. If it is not necessary to change the

setup, press No or

.

NOTE

When no operation is performed for 10 minutes or 60 minutes (depending on the setting of “Auto

Return”), the display except Trend display automatically changes to the Monitor display (or to the

Main display when the MONITOR display is disabled).

NOTE

The display may show “---- (bar)” when starting measurement, which means the sensor is not

ready to measure. But the temperature is displayed normally. Wait until measurement values are

displayed.

While a bar (----) is displayed, as for the mA output, in the case of “Off” or “Low,” [Burn] of the

Output setup (see section 16.3) is fixed 3.6mA, and, in the case of “High,” 22.0mA becomes

fixed.