<7. COMMISSIONING OF SC (Conductivity)>

7-7

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

The output of the selected process parameter is shown as a bar on the bottom of the Main

display or the Home display. And its parameter symbol (for example, Conduct1-TC1 or Diff-

Cond-TC1) is shown above the bar, too. When a selected process parameter is displayed as a

measurement value, the top left number or character is turned to be white number or character

on black background (for example, or

). (Refer to the section 1.2.)

Table 7.2

List of Process Parameters

Sensor

Measurement

Process Parameters

1 or 2

Conductiivity

Conduct1 (2)-TC1

Conduct1 - TC1 :

Conductivity of sensor 1
Temperature compensation 1

Select the temperature conpemsation

method is section 7.1.4.

Temperature1 (2)

Conduct1 (2)-TC2

Resistivity

Resistivity1 (2)-TC1

Temperature1 (2)

Resistivity1 (2)-TC2

Concentration

Conduct1 (2)-TC1

Temperature1 (2)

Concent1 (2)-TC2

Conduct. + Concentr.

Conduct1 (2)-TC1

Temperature1 (2)

Concent1 (2)-TC2

2

Conductiivity

Resistivity

Concentration

Conduct. + Concentr.

Calculated

Redundant

Calculated data and Redundant system are available when two modules are installed on

the instrument. And, these functions are available when the “Conductivity” (or the parameter

including conductivity) or the “Resistivity” is selected on the “Measurement” type setting. The

parameter for the 1st module and the 2nd module should be the same. For wrong selection, an

error is given.
When the process parameter is set at “Calculated” of “Redundant”, don’t change the

“Measurement” type. If it is changed, the process parameter will go back to the default.

<Calculated>
For details, see section 7.7.
<Redundant >
On the Redundant system, when a sensor (Sensor 1) of the 1st module fails, the output is

automatically switched to the output of the 2nd module.
After repairing the Sensor 1, manual reset of redundant system is necessary to return to the

output of 1st module from the output of the 2nd module.

SC

7