<2. WIRING AND INSTALLATION>

2-5

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

CAUTION

The FLXA21 is used with a DC power supply. Do not use an alternating current or 100 volt mains

power supply.

The cable leading to the distributor (power supply) supplies the power to and transmits the output

signal from the FLXA21.
Use a two conductor shielded cable with a minimum cross sectional area of 1.25 mm

2

and an

outside diameter of 6 to 12 mm. The cable gland supplied with the instrument is compatible with

these diameters.
Terminal screw size is M4, and torque of screw up is 1.2 N•m.
Pin terminal, ring terminal and spade terminal can be used for the power supply terminals in the

plastic housing and the stainless housing, and the grounding terminal in the plastic housing. For

the grounding terminal on the stainless steel case, ring terminal should be used.

Pin terminal: pin diameter: max. 1.9 mm

Ring and spade terminal: width: max. 7.8 mm

CAUTION

Cables that withstand temperatures of at least 60°C should be used for wiring.

CAUTION

Always place the shield cover over the power supply and contact terminals for safety reasons

and to avoid interference.

2.4.1 Grounding

The way of connecting the grounding terminal varies depending on whether the housing is plastic

or stainless steel.

For the safety of the user and to protect the instrument against interference, the converter

housing must always be connected to a class D ground (a ground resistance of 100 ohm or less).

CAUTION

The minimum cross sectional area of the protective grounding cable should be 0.75 mm

2

.

l

Grounding the plastic housing

Connect a single separate cable for grounding to the terminal block in the housing (Figure 2.7A).
Use a cable with an outer diameter of 3.4 to 7 mm.
Since the cable is very thin, set the sleeve inside the standard cap of the cable gland and pass

the cable through it. Sleeves are supplied with cable glands (Figure 2.4).
When connecting the converter to a distributor, connect the shield of a double-core shielded

cable to the terminal on the terminal block in the converter. Do not connect the shield to the

ground terminal in the distributor.