5 trend graphics, Trend graphics -10 – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 90

Advertising
background image

<6. OPERATION OF SC (Conductivity)>

6-10

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

Read Logbook:

logbook1-1

Memorandum:

Item
Sensor washed by hand
Enter?
No

logbook1-1 1/1

Memorandum:

Item
Sensor replaced
Enter?
No

2010/02/15 17:04
Sensor replaced
2010/02/15 15:15
Power on

Finish

2010/02/15 15:15
Power on

Finish

logbook1-1 1/1

** When storing message manually.

**

(Fig.6.6)

Sensor washed by hand
Module replaced
Sensor replaced

logbook1-1
logbook1-2

No
Yes

Figure 6.8

Detail display (continued)

6.5

Trend graphics

Pressing

on the Zoom display changes the display to a graphical mode in which the

average measured value is shown on a time scale. The “Live” value is also digitally displayed in

a text box. The time scale (X-axis) and the primary value scale (Y-axis) are set in the “DISPLAY

SETUP” menu (Section 7.7.2).
The screen displays the trend of up to 41 averages of the measurement for each time interval.

The FLXA21 samples the measurements every second. The trending graphic also shows the

maximum and minimum measured values in that interval.
For example, if the time scale is set to 4 hours, then the trend is shown for 4 hours prior to the

actual measurement. Each point on the trend line represents the average over 4Ч60Ч60/41 =

351 measurements (seconds).

NOTE

Updating the trend screen setup resets the current trend graph and starts a new one.

Advertising