4 regulatory compliance, Regulatory compliance -9, 1ph sc isc do – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 21: Sencom

Advertising
background image

<1. INTRODUCTION AND GENERAL DESCRIPTION>

1-9

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

1.4

Regulatory Compliance

Safety: EN61010-1

UL 61010-1

CAN/CSA C22.2 No.61010-1

EMC: EN61326-1 Class A, Table 2 (For use in industrial locations)

EN61326-2-3

AS/NZS CISPR11

Korea Electromagnetic Conformity Standard

Installation altitude:

2000 m or less

Category based on IEC 61010: I (Note 1)

Pollution degree based on IEC 61010: 2 (Note 2)

Note 1: Installation category, called over-voltage category, specifies impulse withstand voltage.

Equipment with “Category I” (ex. two-wire transmitter) is used for connection to circuits in which measures are taken

to limit transient over-voltages to an appropriately low level.

Note 2: Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce

dielectric strength. Degree 2 is the normal indoor environment.

Explosion-proof (Intrinsically safe type and non-incendive) (for suffix code: -EA):

ATEX Intrinsically safe approval

Applicable standard

Explosive Atmospheres

EN 60079-0:2009

General requirements

EN 60079-11:2007

Equipment protection by intrinsic safety “i”

EN 60079-26:2007

Equipment with equipment protection level (EPL) Ga

EN 60529:1992

Degrees of protectionprovided by enclosures (IP Code)

Type of protection

II 1G Ex ia IIC Ga

Group:

II

Category:

1G

T4: for ambient temperature:–20 to 55ºC

T6: for ambient temperature:–20 to 40ºC

Atmosphere pressure: 80 kPa (0.8 bar) to 110 kPa (1.1 bar)

Degree of Protection of the Enclosure: IP66

IECEx Intrinsically safe

Applicable standard

IEC 60079-0:2007

Part 0: General requirements

IEC 60079-11: 2006 Part 11: Equipment protection by intrinsic safety “i”

IEC 60079-26: 2006 Part 26: Construction, test and marking of Group II Zone 0

electrical

apparatus

IEC 60529: 2001

Degrees of protection provided by enclosures (IP Code)

Type of protection

Ex ia IIC Ga

T4: for ambient temperature:–20 to 55ºC

T6: for ambient temperature:–20 to 40ºC

Atmosphere pressure: 80 kPa (0.8 bar) to 110 kPa (1.1 bar)

Degree of Protection of the Enclosure: IP66

FM Intrinsically safe and nonincentive approval

Applicable standard

FM-3600: 2011 Approval Standard for Electric Equipment for use in Hazardous

(Classified) Locations General Requirement

FM-3610: 2010 Approval Standard for Intrinsically Safe Apparatus and Associated

Apparatus for Use in Class I, II, and III, Division 1, Hazardous

(Classified) Locations

FM-3611: 2004 Nonincendive Electrical Equipment for Use in Class I and II, Division 2

and Class III, Divisions 1 and 2, Hazardous (Classified) Locations

FM-3810: 2005 Electrical Equipment for Measurement, Control and Laboratory Use

NEMA 250:1991 Enclosures for Electrical Equipment (1000 Volts Maximum)

ANSI/IEC 60529:2004

Degrees of protection provided by enclosures (IP Code)

ANSI/ISA 60079-0 2009

Part 0: General Requirements

ANSI/ISA 60079-11 2011

Part 11: Equipment protection by intrinsic safety “i”

1

PH

SC

ISC

DO

SENCOM

Advertising
This manual is related to the following products: