<14. CALIBRATION OF DO (Dissolved Oxygen)>

14-2

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

14.1 Air calibration

Span calibration in the ambient atmosphere is the most common and easiest method.
Although setting the Zero Calibration to “Enabled” allows 2-point calibration of Zero (0%) and

Span (100%), only span calibration is generally performed. For zero calibration, see the next

section.

Set the sensor in the maintenance mode. Wash off any stain on the membrane and use a soft

tissue to wipe off any remaining water from the membrane.
Leave the sensor in an environment without temperature fluctuation or air flow for about 10

minutes, such as in an empty container. After confirming that the reading is stable, calibrate the

sensor.

14.2 Water calibration

Water calibration is a method of calibrating dissolved oxygen analyzers in a laboratory. Accurate

results can be obtained if the span calibration is performed carefully in air-saturated water.
Although setting the Zero Calibration to “Enabled” allows 2-point calibration of Zero (0%) and

Span (100%), the reading must have been stable before performing zero calibration. Therefore,

zero calibration takes a relatively long time.
Conduct water calibration in fresh water. For zero calibration in salt water, perform it by manual

calibration.
Set the sensor in the maintenance mode. Wash off any stain on the membrane and use a soft

tissue to wipe off any remaining water from the membrane.

l

Air-saturated water

Pour water (without salt content) into a beaker or other container, and place it on a magnetic

stirrer or other agitator. Put a stirring bar in it and agitate the water, and send air from a pump

(bubbling). It takes 15 to 30 minutes until the water is fully saturated. Then put the sensor in this

water which is still agitated and bubbling, and calibrate it. Keep the sensor at least 3 cm above

the bottom of the beaker to prevent the membrane from directly contacting the air bubbles.

l

Water for zero calibration

Zero calibration takes a long time and normally there is no need to perform it. A clean sensor

does not have a zero current and thus does not require zero calibration.
For the zero calibration water, dissolve 20 to 30 grams of sodium sulfite in 1 liter of desalinated

water.

NOTE

Dissolve sodium sulfite in the water and wait until the reading becomes stable before calibrating

the sensor. Even a healthy sensor needs 40 to 50 minutes. Leave the sensor in the solution.