5 logbook configuration, Logbook configuration -12, Caution – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 195

Advertising
background image

<16. COMMISSIONING OF SENCOM pH/ORP>

16-12

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

Table 16.3

Error configuration

Display item

Description

Default

pH too high

The pH value exceeds 16.00.

Warn.

pH too low

The pH value is lower than –2.00.

Warn.

Temperature too high

Measured process temperature is higher than the

maximum limit.

Warn.

Temperature too low

Measured process temperature is lower than the

minimum limit.

Warn.

ORP too high

The ORP value is higher than 1500 mV.

Off

ORP too low

The ORP value is lower than –1500 mV.

Off

rH too high

The rH value is higher than 100.

Off

rH too low

The rH value is lower than 0.

Off

Matrix config. error

The temperature compensation matrix is not set

properly (see Section 16.2.3).

Fault

Calib. time exceeded

Calibration time exceeds the calibration interval

(see Section 16.2.4).

Off

Wash half-time error

The response to wash is abnormal.

Off

Impedance 1 too high

The sensor must be checked.

Off

Impedance 1 too low
Impedance 2 too high
Impedance 2 too low
SENCOM changed

The SENCOM sensor has been changed.

This error setting can be set to either Off or Warn..

Warn.

CAUTION

If canceling an error configuration could be risky, do not cancel it as a dangerous situation may

result.

NOTE

The “Impedance 1 too high/too low” and “Impedance 2 too high/too low” error items are displayed

in the Error configuration screen. In general, for a pH sensor, “Impedance 1” means “glass

electrode,” and for an ORP sensor, it means “metal electrode.” “Impedance 2” represents the

reference electrode in all cases.

NOTE

Before using sensors which do not have any temperature measuring elements, set both of

“Temperature too high” and “Temperature too low” to Off on the “Error setting” screen. Due to the

open input of temperature signal, a false error may happen as if temperature goes over a high

limit or a low limit.

16.5 Logbook configuration

In “Logbook configuration,” the user configures information to be saved to a logbook or initializes

the logbooks.
Logbooks are used to keep an electronic record of events such as error messages, calibrations,

and programmed data changes. By referring to this log, users can, for instance, easily determine

maintenance or replacement schedules.

Advertising
This manual is related to the following products: