<16. COMMISSIONING OF SENCOM pH/ORP>

16-13

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

In “Logbook configuration,” the user can select “Off,” “1-1,” or “1-2” for each item of interest to be

logged.
The SENCOM logbook cannot be specified.
This can be done for items displayed on the Settings logbook 1/3 to 3/3 screens. Assigning 1-1 or

1-2 to each item allows information to be organized and logged in a logbook.

NOTE

Some events such as power-on are saved into the logbook “1-1”. This logbook may be full earlier.

It is recommended that important information be saved into the logbook “1-2”.

For “Erase logbook”, a specified logbook “1-1”, “1-2” or “SENCOM” can be erased individually.
When the “Warn if logbook full” is set to “Yes”, a warning is given when the logbook come to near

full (up to 3 pages for the SENCOM logbook, up to 13 pages for others).

NOTE

When the logbook gets full, the oldest information is erased automatically.

16.6 Advanced setup

Advanced setup is used to set functions relating to matters other than measurements such as the

selection of settings, tag setting, password setting for protecting calibration and commissioning

operations, date setting, and communication setting.
(“Factory setup” is for service engineers only; there is no item to be set by the user.)

16.6.1 Settings

In “Settings,” select an item to be set as the default value from among “No action,” “Load factory

settings,” “Save user settings,” and “Load user settings.”
When the default values are loaded, the instrument will be restarted. (In the case of “Save user

settings,” it will not be restarted.)
The following parameters are not included in the defaults:

• Tag
• The contents of all logbooks

NOTE

When the “Save user setting” is only selected, save of the user-set parameters will start at once.

To avoid wrong selection, operation in this Settings must be done by

, not by touching the

menu message.

If you select “Load factory settings,” the instrument will be set to the default settings at factory

shipment.
When this item is selected, a screen prompting whether to restart is displayed. If this is no

problem, press “Yes.” Then the “Loading …” message appears and blinks and loading is started.

When the factory settings have been loaded, the instrument will be restarted.

16

SENCOM