<16. COMMISSIONING OF SENCOM pH/ORP>

16-16

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

Wash time/measure time:

Set the wash time in minutes. In a continuous wash, wash time is

replaced by measure time.

Recovery time: Set the recovery time in minutes.
Manual wash:

Select “Disabled” or “Enabled.”

When this item is enabled, a wash cycle can be activated manually. On the

Calibration/Wash screen, press the “Start manual wash cycle” to perform a

manual wash.

Imp2 wash:

Select “Disabled” or “Enabled.”

When this item is enabled, a wash cycle can be started if the “Impedance 2

too high” error occurs with respect to the reference electrode.

Continuous wash: Select “Disabled” or “Enabled.”

A continuous wash cycle is started at the instant when this item is enabled.

In a continuous wash, the “Measure time” and “Interval time” are reversed
(see Figure 16.4).

Tw

Tr

Tint

Tw

Tr

Tint

Tw

Tr

Tint

Tint

Tw’

Tw

Tr

Tw’

F050605_2.ai

Tint: Interval time

Tw: Wash time

Tr: Recovery time

Tint: Interval time

Tw: Measure time

Tw’: Wash time

Tr: Recovery time

When a continuous wash is disabled

When a continuous wash is enabled

Figure 16.4

The instrument has a function for checking whether measurements are normally made after

wash, allowing you to check a sensor response.
The wash recovery check is performed by making use of the time to recover half the wash-cycle

pH change. The half the wash-cycle pH change recovery time replaces the relaxation time by the

recovery time.
Whether to conduct the wash recovery check can be set on the Error settings 2/3 screen. Half

the wash-cycle pH change refers to half the value (1/2∆pH) of the difference (∆pH) between the

pH value during normal measurement of an example in Figure 16.5 and the pH value detected

during wash. If the pH value during wash is greater than 1/2∆pH when 1/3 tR has elapsed, the

electrode is assumed to be good. If it is smaller than 1/2∆pH, then the electrode is assumed to be

bad.
However, for applications where the pH during wash shows almost the same value as the normal

pH, the difference will be approximately zero. In such a case, disable the wash recovery check.

An example of such applications is the monitoring of waste water pH. If you select water jet

cleaning in such a case, the normal pH as well as the washing time pH will be around pH 7; the

difference will be approximately zero, so the recovery time check will not work normally.
This function is available either for ORP measurement only.