<4. COMMISSIONING OF pH/ORP>

4-3

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

4.1

Sensor setup

“Sensor type” setup is determined by the sensor to be connected to the instrument. Select one of

the following three sensor types:

pH:

Only pH is measured. The glass electrode (input 1) is connected to terminal 15

and the reference electrode (input 2) is connected to terminal 13.

ORP:

Only Redox is measured. The metal electrode (input 1) is connected to terminal 15

and the reference electrode (or glass) (input 2) is connected to terminal 13.

pH + ORP: Both pH and ORP are measured simultaneously. The glass electrode is connected

to terminal 15, the reference electrode to terminal 13, and the metal electrode to

terminal 14. This setup also allows rH to be measured.

For best results, connect a liquid earth to terminal 14. For details of the wiring connections,

see Section 2.5.1. If there is no liquid earth, the sensor diagnostic function is disabled; set all of

“Impedance 1 too high/low” and “Impedance 2 too high/low” in “Error configuration” to Off so that

no related error is caused.

NOTE

The selection of “Sensor type” determines the menu structure throughout the instrument.
When pH + ORP is specified as the sensor type, the object under measurement will be displayed

at the top of “Measurement setup.”
If the sensor type is changed from pH to ORP, part of the relevant setup is reset, requiring

reconfiguration.
When ORP is specified, and when an ORP sensor doesn’t have any temperature element,

set both of “Temperature too high/low” in “Error configuration” to Off so that no related error is

caused. (Refer to the section 4.4.) Without temperature input, a temperature shown on the Main

display or the Home display is the maximum or the minimum temperature of the temperature

element’s range set on the Temperature settings. Setting in “Display setup” can make

temperature value on the display not to be displayed. (Refer to the section 4.7.)

4.2

Measurement setup

This section describes how to set up various parameters relating to measurements.
Measurements are performed based on the measurement parameter setup.

For measurement setup, part of the menu structure of settable items changes depending on the

“Sensor type” selected in “4.1 Sensor setup.”

4.2.1 Measurement

If “pH” or “ORP” is selected in “Sensor type,” a measurement type does not need to be selected

and it is not displayed.
If “pH + ORP” is selected in “Sensor type,” five measurement types can be selected for the

“Measurement” item at the top of the Measurement setup screen. Select a suitable measurement

type for the application.
Moreover, associated items are added to parameters in “Error configuration,” “Display setup,”

etc.; check the setting condition of each item.

PH

4