<7. COMMISSIONING OF SC (Conductivity)>

7-12

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

7.5.3 Passwords

Calibration and commissioning operations can be separately protected by each password. To

protect execute operations, enter a password in Execute’s input field. To protect commissioning

operations, enter a password in Commissioning’s input field. By default, both input fields are

empty. When a password input field is empty, operation is not password-protected. A password

can contain up to 8 characters.

When you set a password, always take a note of it.

When a password is set, input of the password is necessary to enter the password-protected

operation. After inputting the password, the display will change to an operator ID input display.

When an operator ID is input, its operation is recorded into a logbook. The input of an operator ID

is not necessary to enter the operation. An operator ID can contain up to 4 characters.

7.5.4 Date/Time

The Logbooks and trend graphs use the clock/calendar as a reference. The current date and

time is set here. The time display format can be selected from among three types.

7.5.5 Communication

In “Communication,” select the communication setting from among None, HART, and PH201G.

The burn down current value is 3.6 mA.

NOTE

To make the change of “Communication” valid, turn off the power supply once, and reboot.

In the case of “None,” there is not the problem even if you do not change it as “HART” of default

value.

n

HART

Select this menu when HART communication is made.
In the HART setup screen, specify the network address and set up parameters for SV, TV, and

FV.
(PV is linked with the “process parameter” setting in “Output settings” and cannot be changed

here.)

l

Network address

For 1-to-1 communication, leave the default value [0] unchanged. For multi-drop where multiple

HART devices are connected on a bus, set addresses in 1 to 15. In this case, the mA output will

be fixed to 4 mA.

l

PV

PV is a parameter selected for analog output; it cannot be changed here.

l

SV, TV, FV

The SV, TV, and FV parameters are items that the user must set up. Selectable items differ

depending on the sensor type in “Sensor setup” and settings in “Measurement setup.”
If blank is selected for a parameter, items below that parameter must all be set to blank. If an item

is blank, those below it cannot be set to a status other than blank.