<3. OPERATION OF pH/ORP>

3-9

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

As shown in Figure 3.8, the date is predicted based on the intersection point of the upper or lower

limits and the extrapolated line of the values obtained by the least squares method.

Day

Value

Lower limit

Upper limit

Current

date

Projected

maintenance date

F0310.ai

Figure 3.8

The status shows the certainty of the projected maintenance date in terms of the correlation

coefficient R. Tables 3.1 and 3.2 show respective display patterns.

Table 3.1

Display pattern of the projected maintenance date

Projected date

- -

0-1 month

1-3 months

3-6 months 6-12 months Over 1 year

- - :

cannot be predictable due to insufficient data

Table 3.2

Display pattern of the status

Status

(- - - - -)

(R < 0.50)

(Poor)

(0.50 ≤ R < 0.70)

(Reasonable)

(0.70 ≤ R < 0.85)

(Excellent)

(0.85 ≤ R < 1.00)

n

Projected replacement

The projected replacement function predicts the date when the sensor will need replacing for

maintaining the measurement accuracy, based on the pH zero and pH slope on each calibration,

and reference impedance (input-2 impedance) after each calibration. The projected replacement

date is predicted based on these parameters stored upon calibration, and displayed the same as

that of projected maintenance. For details, see the description about the projected maintenance.

Since three parameters (pH zero, pH slope, and input-2 impedance after calibration) are used for

this projection, the nearest coming day is selected as the projected replacement date from the

extrapolated line of the values obtained by the least squares method.

n

PH module (sensor)

With this screen, you can check the module productions number and software revision of the

installed module.

n

HOUSING ASSY

With this screen, you can check the module productions number, software revision, and HART

device revision of the housing assembly.

n

Read logbook

The FLXA21 has two types of logbook per sensor to store history information on events, such as

changed settings and calibrations.
By selecting one of the logbooks that you wish to check, you can retrieve and check this

information. Storage of history information on each event in a logbook or which logbook to use for

storage can be set up on the Configure logbook screen. For details, see section 4.5.