2.1

Installation site

The FLXA21 is weatherproof and can be installed both inside and outside. It should, however,

be installed as close as possible to the sensor to avoid long cable runs between the instrument

and sensor. When the pH sensor is used, the cable length including the sensor cable should not

exceed 20 meters (65.6 feet). 60 meters when using BA10 extension box and WF10 cable.

Select an installation site where the ambient temperature and humidity are within the limits of the

instrument specifications (see chapter 2). If the instrument is installed outside and exposed to

direct sunlight, a sun shade hood should be used.

Ambient Operating Temperature: -20 to +55 ºC
Storage Temperature:

-30 to +70 ºC

Humidity:

10 to 95% RH (Non-condensing)

Select an installation site that meets the following conditions.

• Mechanical vibrations and shocks are negligible
• No relay switch and power switch are installed close to the converter
• There is space for cable connection beneath the cable glands
• Not exposed to direct sunlight or severe weather conditions
• Maintenance is possible
• No corrosive atmosphere

2.2

Removing the wiring cover

Before installing this instrument, open the front panel and remove the wiring covers, and then

install the cable glands.
Remove all the wiring covers. They are fixed at the places marked with ○ (circle) and Δ (triangle)

in Figure 2.2. Be careful not to damage the pawls when removing the covers.

NOTE

Be careful not to lose the wiring cover screws.

Wiring cover 1

(for the first module)

Wiring cover 2

(for the second module)

Wiring cover 3

(for the power supply and grounding)

Figure 2.2

Wiring cover