<12. OPERATION OF DO (Dissolved Oxygen)>

12-8

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

n

DO module (sensor)

With this screen, you can check the module productions number and software revision of the

installed module.

n

HOUSING ASSY

With this screen, you can check the module productions number, software revision, and HART

device revision of the housing assembly.

n

Read logbook

The FLXA21 has two types of logbook per sensor to store history information on events, such as

changed settings and calibrations.
By selecting one of the logbooks that you wish to check, you can retrieve and check this

information. Storage of history information on each event in a logbook or which logbook to use for

storage can be set up on the Configure logbook screen. For details, see section 13.5.
History information on events are automatically stored on the preset conditions.
In addition to this storing, following three messages can be manually stored in the logbook;

"Sensor washed by hand", "Module replaced", "Sensor replaced"

To store these messages, press

and select one of three messages from the Item on the

Memorandum screen. Its event date/time will be the time when a message is selected and

entered.

When a password for Commissioning is set on the passwords’ setup, pressing

requires

entering the password. (Refer to the section 13.6.3.)

Read Logbook:

logbook1-1

Memorandum:

Item
Sensor washed by hand
Enter?
No

logbook1-1 1/1

Memorandum:

Item
Sensor replaced
Enter?
No

2010/02/15 17:04
Sensor replaced
2010/02/15 15:15
Power on

Finish

2010/02/15 15:15
Power on

Finish

logbook1-1 1/1

** When storing message manually.

**

(Fig.12.6)

Sensor washed by hand
Module replaced
Sensor replaced

logbook1-1
logbook1-2

No
Yes

Figure 12.7

Detail display (continued)

Advertising