<13. COMMISSIONING OF DO (Dissolved Oxygen)>

13-11

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

If blank is selected for a parameter, items below that parameter must all be set to blank. If an item

is blank, those below it cannot be set to a status other than blank.

For more information on HART communication, see the Technical Information (TI 12A01A02-

60E) in the attached CD-ROM.

n

PH201G

Select this menu if the PH201G distributor is connected to the instrument.

In the PH201G setup screen, make settings for “Hold contact,” “Fail contact,” and “Wash

contact.”

l

Hold contact

Select Disabled or Enabled.
When this item is enabled, the output will be held according to the setting of “Hold type” on the

Hold setup screen.

l

Fail contact

Select a status from among “Fail + Warn,” “Fail only,” and “Disabled.”
This setting depends on the error configuration. See “13.4 Error configuration.”
“Fail” corresponds to “Fault.”

l

Wash contact

Select Disabled or Enabled.
When this item is enabled, items for a wash can be set on the Wash settings screen.
In the Wash settings screen, set the interval time, wash time, and recovery time and setup of

various washes.

Interval time:

Set the wash interval in hours.

Wash time/measure time:

Set the wash time in minutes. In a continuous wash, wash time is

replaced by measure time.

Recovery time: Set the recovery time in minutes.
Manual wash:

Select “Disabled” or “Enabled.”

When this item is enabled, a wash cycle can be activated manually. On the

Calibration/Wash screen, press the “Start manual wash cycle” to perform a

manual wash.

Imp2 wash:

Select “Disabled” or “Enabled.”

When this item is enabled, a wash cycle can be started if the “Impedance 2

too high” error occurs with respect to the reference electrode.

Continuous wash: Select “Disabled” or “Enabled.”

A continuous wash cycle is started at the instant when this item is enabled.

In a continuous wash, the “Measure time” and “Interval time” are reversed
(see Figure 13.3).