Mounting methods

See Figure 2.9. The FLXA21 has various mounting possibilities.

• Panel mounting using optional mounting hardware (/PM or /UM)
• Wall mounting using optional mounting hardware (/U or /UM)
• Pipe mounting using optional mounting hardware (/U or /UM) on a horizontal or vertical

pipes (Nominal size: 50A)

The universal mounting kit (/UM) contains the pipe and wall mounting hardware (/U) and the

panel mounting hardware (/PM). Some can be left unused.

195

0

+1

138

0

+1

138

185

Unit: mm

Panel mounting

Wall mounting

Pipe mounting

Note: The attached hardware is the same

as that for the pipe mounting.

Only the bracket is used for the

wall mounting.

Note: Horizontal pipe mounting is

possible by turning the bracket

90 degrees.

M8 bolt (not supplied)

Prepare bolts of the same

length as the mounting hole.

Note: For a sun shade hood (option),

use the upper two bracket

mounting screws to fix it.

Bracket

Wall mounting holes (x3)

Wall

Bracket mounting screws*

Bracket mounting screws*

Fixing screws (x2)

Panel

Spacing panel cutout

Nuts (x2)

Washers (x2)

Stanchion (50A pipe)

Bracket mounting screws*

Bracket

Bracket

Bracket for piping

U-bolt

* Tighten the four screws to a torque of 2 N·m.

* Tighten the four screws to a torque of 2 N·m.

* Tighten the four screws to a torque of 2 N·m.

Figure 2.9

Three types of mounting with the optional mounting hardware

See General Specifications (GS 12A01A02-01E) for the dimensions of the FLXA21 and

mounting hardware.