<17. CALIBRATION OF SENCOM pH/ORP>

17-5

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

17.1.3

Sample calibration

A sample calibration is a single-point calibration for only the zero (asymmetric). It adjusts the

recorded reading to a collected sample value. Press [Take Sample] to record a collected sample

value in memory. Re-enter the Sample Cal. screen and press [Start calibration] to perform a

sample calibration. This updates the recorded data.

NOTE

When a sensor or an electrode is exchanged or replaced, sensor wellness data should be reset.
When a sensor is replaced, the replacement can be recorded manually into a logbook. (Refer to

the figure 15.11.)

17.2 Temperature calibration

For the most accurate measurements, it is important to have a precise temperature

measurement. Measure the temperature with a high-precision thermometer and adjust the

sensor reading accordingly. For best accuracy, this should be done as near to the normal

operating temperature as possible.

17.3 ORP calibration (rH calibration)

The calibration modes for ORP or rH are “Manual” and “Sample”. No automatic calibration

feature is available in an ORP or rH calibration.

Calibration is performed stepwise. Follow the prompts displayed on the screen.
A stability check is made at each measurement point. Proceed to the next step only after the

reading has stabilized.

NOTE

When a sensor or an electrode is exchanged or replaced, sensor wellness data should be reset.
When a sensor is replaced, the replacement can be recorded manually into a logbook. (Refer to

the figure 15.11.)