<10. COMMISSIONING OF ISC (Inductive Conductivity)>

10-6

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

l

Unit for table

Select the concentration display units from among %, ppt, ppm, and ppb. Changing the unit will

not result in a re-calculation of the table. Reenter values in the additional concentration table.

10.1.7

Sensor diagnostic settings

This screen is used to set items relating to sensor diagnostics displayed on the screens invoked

by pressing

.

Gauges are displayed for only parameters that have been enabled in “Sensor diag. settings.”
Parameters set to Disable are provided with a bar display.

The setting parameters include Progress time and Heat cycle. It is also possible to set the “Bad

limits” of the progress time and heat cycle and the “Heat cycle temp” and “Heat cycle time” of the

heat cycle.

10.2 Output setup

The general procedure is to first define the function of the output, Output or Simulate. Then, set

the process parameters associated with the output. On the Output, an output of measured value

is selected. On the Simulate, a simulation value can be set.
And, the parameters for HOLD function can be set on this setting.

l

Output

The output signal is a current value specified by the following parameters.

Process parameter
Available process parameters depend on the selected “Measurement” item in Measurement

setup. Refer to Table 10.2.
The output of the selected process parameter is shown as a bar on the bottom of the Main

display. And its parameter symbol (for example, Conduct1-TC1) is shown above the bar, too.

When a selected process parameter is displayed as a measurement value, the top left number

or character is turned to be white number or character on black background (for example, ).

(Refer to the section 1.2.)

Table 10.2

List of Process Parameters

Sensor

Measurement

Process Parameters

1

Conductiivity

Conduct1-TC1

Conduct1 - TC1 :

Conductivity of sensor 1
Temperature compensation 1

Select the temperature conpemsation

method is section 10.1.4.

Temperature1

Conduct1-TC2

Concentration

Conduct1-TC1

Temperature1

Concent1-TC2

Conduct. + Concentr.

Conduct1-TC1

Temperature1

Concent1-TC2

Setup
Select one of the output methods: Linear and Table.

Linear:

Set the 0% and 100% values.

Table:

This allows the configuration of an output curve by 21 points (5% intervals).

(The 0% and 100% values must be entered.)