3 home display, main display and monitor display, Home display, main display and monitor display -4, Nconfigure sensor – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 84: Ntemperature settings, Nma (output)

Advertising
background image

<6. OPERATION OF SC (Conductivity)>

6-4

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

n

Configure sensor

From among the sensor types displayed on this screen, select an appropriate electrode for the

sensor used. The measurement units can also be selected from among “/cm” and “/m”.
The cell constant (factory default) is determined by factory calibration made during

manufacturing.
The cell constant is indicated on the sensor. If a new sensor is used, the cell constant indicated

here should be changed. When this value is changed, the real cell constant will also be changed.
For details, see section 7.1.2.

n

Temperature settings

Select a suitable temperature element from among those displayed and set it up.
Celsius (ºC) or Fahrenheit (ºF) temperature scale can be selected.
For details, see section 7.1.3.

n

mA (output)

Select an appropriate process parameter from among those displayed and set it. The mA output

has been set to 0 – 500 µS/cm or 0 to 20 Ω cm at factory shipment. If a high resolution is required

in a consistent measurement process, set this parameter to a value suitable for the process.
For details, see section 7.2.

6.3

Home display, Main display and Monitor

display

Pressing

changes the screen to the Main display shown in Figure 6.3 (or the Home display

shown in Figure 6.4).

If two sensors are connected, pressing

on the Main display changes the display to the

Home display shown in Figure 6.4.

If only one sensor is connected,

is grayed out and disabled on the Main display.

10.38

Tag:FLXA21–SC

25.0

4mA

20mA

Conduct1-TC1

°C

mS/cm

Figure 6.3

Example of main display

6.35

Tag:FLXA21–SC

24.9

4mA

20mA

Conduct1-TC1

°C

mS/cm

Tag:FLXA21–SC

25.0 °C

mS/cm

10.38

Figure 6.4

Example of home display

Advertising
This manual is related to the following products: