1 measurement setup, 1 measurement, 2 configure sensor – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 95: 3 temperature settings, Measurement setup -3 7.1.1 measurement -3 7.1.2, Configure sensor -3, Temperature settings -3, Sc 7

Advertising
background image

<7. COMMISSIONING OF SC (Conductivity)>

7-3

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

7.1

Measurement setup

This section describes how to set up various parameters relating to measurements.
Measurements are performed based on the measurement parameter setup.

7.1.1 Measurement

Select a measurement parameter from Conductivity, Resistivity, Concentration, and Conduct.

+ Concentr. The setting of the measurement parameter changes the menu structure in Error

configuration, Display setup, etc. accordingly.

7.1.2

Configure sensor

This section describes how to configure the sensor to be connected.

l

Sensor type

Choose the sensor type from “2-electrode” and “4-electrode.”
Two-electrode sensors are usually used for conductivity and/or resistivity measurements.
However, polarization of the electrodes may occur at high conductivity ranges, which may cause

an error in conductivity measurement. To avoid this, use 4-electrode sensors.

l

Measuring unit

Either /cm or /m can be chosen here.
The process values will be expressed in S/cm or S/m in conductivity mode or Ω.cm or Ω.m in

resistivity mode, respectively.

l

Cell constant (factory)

The cell constant given at factory shipment is set.
When a new sensor is connected, set the new cell constant which is indicated on the sensor.
When a cell constant is changed, a calibrated cell constant (c.c. (adjusted)) will be the same

value of the new cell constant (factory) after resetting an old one.
The cell constant is indicated on the sensor nameplate, etc. by either of the following two ways:

• Cell constant indication and input methods

(1) When the cell constant itself is written on the sensor nameplate, enter it directly. (for

electrodes SC211G, SC8SG, and SC4AJ)

(2) When the “percentage of a deviation from the nominal cell constant (±□.□%)” is indicated on

the sensor nameplate, enter the cell constant as follows. (for electrode SC210G)

In case of the indication of a nominal cell constant of 5 cm

-1

and a deviation value of

(CORR.%=-1.1):

The input cell constant is 5 x (100 – 1.1) / 100 = 4.945; enter this 4.945.

7.1.3

Temperature settings

Select the temperature element used for compensation from among Pt1000, Pt100, Ni100, 8k55,

and Pb36 (JIS 6k). Select the same type as the temperature element that is actually connected.
If a thermistor is used for the temperature sensor, select Pb36 (JIS 6k).

Celsius (°C) or Fahrenheit (°F) temperature units are available. If the unit is changed, the

following values are also recalculated automatically to the new unit:

• Manual temp.

• Reference temp.

• Temp. coefficient

• Temp. ranges in the matrix

SC

7

Advertising
This manual is related to the following products: