<7. COMMISSIONING OF SC (Conductivity)>

7-6

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

l

Concentration table

Concentration can be calculated by using the temperature compensation matrix or the additional

concentration table.

• By using the temperature compensation matrix

Set the “Additional table” to “Disabled”. The concentration can be obtained from the

temperature compensation matrix (based on the relation between the conductivity at the

reference temperature and the concentration).

Select Temp. compensation → Method, and then select Matrix. Next, choose a solution or

User defined 1 or 2 (see Section 7.1.4). The concentration cannot be obtained by any other

temperature compensation method (TC or NaCl).

• By using the additional concentration table

The additional concentration table indicates the relation between the conductivity and the

concentration at the reference temperature. Set the “Additional table” to “Enabled”. The

concentration can be obtained from the additional concentration table.

After entering values, select “Yes” for “Check values?”. Values are interpolated into empty

cells. If any mandatory cells are left blank, an error will be issued. An error message is also

issued if there is any error in the concentration table.

After completing entering values in the additional concentration table, set “Additional table”

to “Enabled”.

l

Unit for table

Select the concentration display units from among %, ppt, ppm, and ppb.
Changing the unit will not result in a re-calculation of the table. Reenter values in the additional

concentration table.

7.1.7

Sensor diagnostic settings

This screen is used to set items relating to sensor diagnostics displayed on the screens invoked

by pressing

.

Gauges are displayed for only parameters that have been enabled in “Sensor diag. settings.”
Parameters set to Disable are provided with a bar display.

The setting parameters include Progress time and Heat cycle. It is also possible to set the “Bad

limits” of the progress time and heat cycle and the “Heat cycle temp” and “Heat cycle time” of the

heat cycle.

7.2

Output setup

The general procedure is to first define the function of the output, Output or Simulate. Then, set

the process parameters associated with the output. On the Output, an output of measured value

is selected. On the Simulate, a simulation value can be set.
And, the parameters for HOLD function can be set on this setting.

l

Output

The output signal is a current value specified by the following parameters.

Process parameter
Available process parameters depend on the selected “Measurement” item in Measurement

setup. Refer to Table 7.2.