<4. COMMISSIONING OF pH/ORP>

4-12

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

Table 4.2

Error configuration

Display item

Description

Default

pH too high

The pH value exceeds 16.00.

Warn.

pH too low

The pH value is lower than –2.00.

Warn.

Temperature too high

Measured process temperature is higher than the

maximum limit.

Warn.

Temperature too low

Measured process temperature is lower than the

minimum limit.

Warn.

ORP too high

The ORP value is higher than 1500 mV.

Off

ORP too low

The ORP value is lower than –1500 mV.

Off

rH too high

The rH value is higher than 100.

Off

rH too low

The rH value is lower than 0.

Off

Matrix config. error

The temperature compensation matrix is not set

properly (see Section 4.2.3).

Fault

Calib. time exceeded

Calibration time exceeds the calibration interval

(see Section 4.2.4).

Off

Wash half-time error

The response to wash is abnormal.

Off

Impedance 1 too high

The sensor must be checked.

Off

Impedance 1 too low
Impedance 2 too high
Impedance 2 too low

CAUTION

If canceling an error configuration could be risky, do not cancel it as a dangerous situation may

result.

NOTE

The “Impedance 1 too high/too low” and “Impedance 2 too high/too low” error items are displayed

in the Error configuration screen. In general, for a pH sensor, “Impedance 1” means “glass

electrode,” and for an ORP sensor, it means “metal electrode.” “Impedance 2” represents the

reference electrode in all cases. The “Impedance 1 too high/too low” and “Impedance 2 too high/

too low” errors are applied to sensors with a liquid earth. Sensors with a liquid earth include

PH8EFP, PH8ERP, PH8EHP, OR8ERG, OR8EFG, etc. For sensors not equipped with a liquid

earth, these errors should all be set to Off; these sensors include HA405, HA406, DPA406,

DPA405, DPAS405, HF405, HA485, DPA485, etc. Moreover, for measuring a solution with low

conductivity such as when using a pH meter for pure water, set the “Impedance 2 too high” error

to Off. Otherwise, no impedance can be detected.

NOTE

When ORP is specified as a sensor type in “Sensor setup”, and when an ORP sensor, such as

OR8ERG and OR8EFG, doesn’t have any temperature element, set both of “Temperature too

high/low” in “Error ” to Off so that no related error is caused. Due to the open input of temperature

signal, a false error may happen as if temperature goes over a high limit or a low limit.