3 error configuration, Error configuration -9, Caution – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 101: Sc 7

Advertising
background image

<7. COMMISSIONING OF SC (Conductivity)>

7-9

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

During the Commissioning or the Quick Setup, the mA output is automatically held. The preset

value depends on a setting on the “Last or fixed”.

“Last”:

The preset value is a value measured just before hold condition.

“Fixed”:

The preset value is a value set in the “Fixed value mA”.

When the “Fixed” is selected, set a mA value in the “Fixed value mA”.

Selection on the “Hold during Calibration” decides to activate or deactivate the hold function

automatically during calibration.

“Enabled”: Activation of the automatic hold function
“Disabled”: No automatic hold function

7.3

Error configuration

In Error configuration, configure the statuses of various error causes.

This allows the system to notify the user of the occurrence of an error according to the status

categories in the Error configuration.
Select a status category from among Off, Warn. (Warning), and Fault.
“Fault” automatically performs burn-out. When Burn has been set to Off (7.2 Output setup), only

the error message is displayed.
“Warn.” displays an error message.
When selecting PH201G in the communication setting, make sure that the “Fail contact” setting

is appropriate.

The settable causes of errors are determined based on the settings of the Sensor setup and

Measurement setup, and a status category is set to the causes displayed in the Errors 1/3 to 3/3

screens.

Table 7.3

Error configuration

Display item

Description

Default

Conductivity too high

(or Concentration)

Conductivity or resistivity is lower than the minimum limit.

Warn.

Conductivity too low

(or Concentration)

Conductivity or resistivity is higher than the maximum limit.

Warn.

Temperature too high

Measured process temperature is higher than the maximum limit.

Warn.

Temperature too low

Measured process temperature is lower than the minimum limit.

Warn.

Polarization detec

Polarization of the electrodes is detected.

Off

Calibr. time exceeded

Calibration time exceeds the calibration interval (see Section 7.1.5).

Off

USP

Conductivity exceeds the USP limit.

Off

Configuration error:

1st comp. matrix

2nd comp. matrix

The temperature compensation matrix is not defined properly (see

Section 7.1.4).

Fault

Concentration table

The concentration table is not defined properly (see Section 7.1.6).

Fault

CAUTION

If canceling an error configuration could be risky, do not cancel it as a dangerous situation may

result.

SC

7