7 sensor diag. settings, 3 output setup, Sensor diag. settings -5 – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 156: Output setup -5, Do 13

Advertising
background image

<13. COMMISSIONING OF DO (Dissolved Oxygen)>

13-5

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

13.2.7

Sensor diag. settings

This screen is used to set items relating to sensor diagnostics displayed on the screens invoked

by pressing

.

Gauges are displayed for only parameters that have been enabled in “Sensor diag. settings.”
Parameters set to Disable are provided with a bar display.

The setting parameters include Progress time and Heat cycle. It is also possible to set the “Bad

limits” of the progress time and heat cycle and the “Heat cycle temp” and “Heat cycle time” of the

heat cycle.

13.3 Output setup

The general procedure is to first define the function of the output, Output or Simulate. Then, set

the process parameters associated with the output. On the Output, an output of measured value

is selected. On the Simulate, a simulation value can be set.
And, the parameters for HOLD function can be set on this setting.

l

Output

The output signal is a current value specified by the following parameters.

Process parameter
For the available process parameters, see Table 13.2.
The output of the selected process parameter is shown as a bar on the bottom of the Main

display or the Home display. And its parameter symbol (for example, Oxygen1 or Diff-Oxygen1)

is shown above the bar, too. When a selected process parameter is displayed as a measurement

value, the top left number or character is turned to be white number or character on black

background (for example, or

). (Refer to the section 1.2.)

Table 13.2

List of Process Parameters

Number of sensors

Process Parameters

1

Oxygen1 (2)
Temperature1 (2)

2

Oxygen1
Temperature1
Oxygen2
Temperature2
Calculated*1
Redundant*2

*1: Refer to section 13.8, Calculated data setup.

*2: Refer to the <Redundant system> below.

Calculated data and Redundant system are available when two modules are installed on the

instrument.

<Redundant system>
On the Redundant system, when a sensor (Sensor 1) of the 1st module fails, the output is

automatically switched to the output of the 2nd module.
After repairing the Sensor 1, manual reset of redundant system is necessary to return to the

output of 1st module from the output of the 2nd module.

DO

13

Advertising
This manual is related to the following products: