1 cell constant (manual), 2 cell constant (automatic), 3 air calibration – Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 User Manual

Page 110: Cell constant (manual) -2, Cell constant (automatic) -2, Air calibration -2

Advertising
background image

<8. CALIBRATION OF SC (Conductivity)>

8-2

IM 12A01A02-01E

5th Edition : Oct. 31, 2013-00

NOTE

When a sensor is exchanged or replaced, sensor wellness data should be reset.
When a sensor is replaced, the replacement can be recorded manually into a logbook. (Refer to

the figure 6.9.)

8.1

Cell constant (manual)

The intention of this calibration routine is to fine-tune a sensor for which only the nominal cell

constant is known, or to recalibrate a sensor that has been changed (or damaged) during

operation. Choose the 1st or 2nd temperature compensation to suit the calibration solution used.

The solution with an appropriate precision should be prepared or purchased. Allow the sensor to

reach stable readings for both temperature and conductivity before adjusting to the value of the

corresponding calibration solution.
Calibration here is performed by measuring a solution with known conductivity at the set

reference temperature. The calibration solution of an accurate concentration should be prepared

by dissolving a certain amount of solute in water. Allow the conductivity to become stable at the

set reference temperature (initial setting of 25°C) of the instrument and obtain the conductivity

from literature data, etc. Appendix 2 shows the conductivity values of sodium chloride (NaCl) as a

typical example.

8.2

Cell constant (automatic)

This routine is built around the test method described in International Recommendation No.

56 of OIML (Organisation Internationale de Metrologie Legale). It allows the direct use of the

solutions prescribed in the test method, automatically selecting the appropriate temperature

compensation. The look-up table is used to find an appropriate conductivity reading for the

measured temperature. See Appendix 2 for OIML solutions.

8.3

Air calibration

It is not necessary to conduct general air calibration unless the cable length is long and a low

conductivity needs to be measured.
With the clean dry cell in open air, the reading should be zero. The air cal compensates for

excess cable capacitance, and gives a better accuracy at low readings. This should be done

when a sensor is installed or replaced. After some time in service, a dirty sensor may well show

a high zero offset because of fouling. Clean the sensor and try again. Moreover, air calibration

should be made in an environment free of electromagnetic interference.

NOTE

The temperature compensation is not performed around zero. In this case, a warning may be

issued. See “● For all temperature compensation” on App. 2-8.

Advertising
This manual is related to the following products: