67 displays – Yokogawa PC-Based MX100 User Manual

Page 18

Advertising
background image

1-11

SM MX100-01E

Principles of Operation

1

2

3

4

5

6

7

Displays

The MW100 indicates its operating conditions with the following displays.
• 7-segment LED
• Status indicators
• Communication status LED

7-Segment LED

Displays the MW100 Data Acquisition Unit’s unit number, operation status, end of
operation, and errors.

• Unit Number Display
Unit numbers can be set from 00 to 89.

-

is displayed.

• Display of the Self-Test Operation on Startup
When the power is turned ON the setting of dip switch 1 is displayed

followed by

the operation preparation status

, and then a self check is performed. While the

self check is in progress, the following displays are repeated.

• Key Lock Status
A key lock function is included for preventing accidental manipulation of the MW100

front panel keys. The key lock status is indicated by a dot at the bottom of the unit
number. The example shown is for a unit of number 00.

Unit number and dot

Unit number

• Keylock release

• Keylock

• In-Progress Display
The following displays cycle while the CF card is being accessed or while calibration

is being performed.

• Operation Error Display
In error Exxx (where xxx is a three-digit number), the code is divided into two parts

which are displayed alternately. In the first part, the letter E appears in the left digit
with the hundreds digit of the error code to the right, and the second part consists of
the last two digits of the error code.
Example: Error code E234

Up to three error codes are saved.
For the contents of error codes and their meanings, see section 4.1, “Errors Displayed

on the 7-Segment LED and Corrective Actions.” You can clear one error that is
displayed by pressing the Stop key.

1.4 Functions of the MW100 Main Module

Advertising
This manual is related to the following products: