2-37

SM MX100-01E

Testing

1

2

3

4

5

6

7

2.8.2

Withstanding Voltage Test

For this test, the Ten-Channel, High-Speed Digital Input Module is installed with the
MX100, MW100 or equivalent device.

Procedure

1.

Measure the withstanding voltage at the same locations indicated in step 1 of the

insulation resistance test. When performing the test, you must fasten the module
being tested to the MX150 base plate jig with screws.

2.

Confirm that the criterion below is met.

Input terminal to earth terminal: AC2300 Vrms for 1 minute, 2 mA or less

2.8.3

Display Input Test

For this test, the Ten-Channel, High-Speed Digital Input Module is installed with the
MX100, MW100 or equivalent device.

Procedure

1.

Allow a warm-up of 10 minutes or more.

2.

Set the measurement interval on the module being tested to 500 ms.

3.

Input DC voltage and resistance to each channel of the Ten-Channel High-Speed

Digital Input Module, and confirm each input value.

If the /NC option is installed, use B8720BF during the test.

(When the test is complete, be sure to remove the B8720BF.)

<MX115-D05-H10>

Syntax

Input Voltage,

Resistance

Measurement

Interval

Judgment Criterion

LEVEL

+1 V or less

500 ms

That the input channel reads 0
That the channels with no input read 1

LEVEL

+3 V or more 500 ms

That the input channel reads 1
That the channels with no input read 0

CONTACT 100 Ω or less 500 ms

That the input channel reads 1
That the channels with no input read 0

CONTACT 100 kΩ or

more

500 ms

That the input channel reads 0
That the channels with no input read 1

<MX115-D24-H10>

Syntax

Input Voltage,

Resistance

Measurement

Interval

Judgment Criterion

LEVEL

+6 V or less

500 ms

That the input channel reads 0
That the channels with no input read 1

LEVEL

+16 V or more 500 ms

That the input channel reads 1
That the channels with no input read 0

CONTACT 100 Ω or less 500 ms

That the input channel reads 1
That the channels with no input read 0

CONTACT 100 kΩ or

more

500 ms

That the input channel reads 0
That the channels with no input read 1

2.8.4

ID Check

This test confirms whether the module and terminal board types are correctly identified
when module information is obtained from the jig on the MX100 or equivalent.

Procedure

Exchange the terminal board from the B8720BF to a screw terminal block or equivalent,
and then check the ID before and after the new board is installed.
When exchanging a terminal board, you must make sure that the power to the MX100 or
equivalent is turned OFF.

2.8 Tests of the 10-CH, High-Speed Digital Input Module (MX115-D05-H10)