2.7.9

Dip Switch Operation Test (-B12, -B35 Only)

You can check whether the dip switches are operating normally by confirming that
measurements can be taken using the 1-gauge method.
(Confirmation using the 4-gauge method can be made with the measurement accuracy
and channel error tests. )

1.

Set the dip switches for the 1-gauge, 2-wire method.

Dip switch settings

ON

1 2 3 4 5

SW1 to SW3: ON
SW4 to SW5: OFF

2.

Connect resistance independantly to all channels as shown in the figure below.

The wiring resistance of the module itself is large, and therefore needs to be
reduced as much as possible.

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15
16

ON

1

2

3

4

5

ON

1

2

3

4

5

ON

1

2

3

4

5

ON

1

2

3

4

5

A/+V
B/ L
C/-V
D/ H

CH1

A/+V
B/ L
C/-V
D/ H

CH2

A/+V
B/ L
C/-V
D/ H

CH3

A/+V
B/ L
C/-V
D/ H

CH4

Resistance

Resistance

Resistance

Resistance

Test module

Resistance stated value

-B35

350Ω ± 0.01%

-B12

120Ω ± 0.01%

Resistance stated value

3.

Restore the initial balance values of all channels to the initial values from a PC via

communication.

4.

Enter the following settings.

Measured CH

All CH

Measurement range

2000 µSTR

Measurement interval

200 ms

Filters

OFF

Duration

10 s or more

5.

Confirm that the measured values are within the stated values.

Measured Value Stated Value

0

µ

STR ±400

µ

STR

2.7 Tests of the 4-CH Strain Input Module