2.4.6

Reference Junction Compensation Accuracy

Cautions When Performing This Test

• The test module uses slots number 0 through 4. A dummy input module (MX110 H04

INPUT MODULE or equivalent) is installed into slot 5.

When performing a test using fewer than 5 modules, install the modules into

contiguous slots starting with slot 0, with the dummy input module installed in the last
slot. The dummy input module is installed simulate common installation conditions for
its neighboring module.

Example: when testing 3 modules at once

Dummy

input

module

MX110

H04

INPUT

#2

MX110

H04

INPUT

#3

MX110

H04

INPUT

#1

MX100

Slot 5

Slot 3

Slot 4

Slot 0

Slot 2

Slot 1

Install the EUT modules to test from the right and then the dummy module last.
Use another MX110-UNV-H04 as the dummy.

• Using a calibrated thermocouple (Type-T), confirm that the criteria are met when

factoring in the calibration value.

• Monitor the 0°C standardized temperature tank, confirming that it stays at 0±0.01°C.
If the temperature tank is off, confirm that the difference is within the allowable range.
• When using a zero-controlled device, the tip of the thermocouple floats 5-10 mm off

the bottom. However, that distance varies depending on the model. Please see the
user’s manual.

• Perform the test in a stable environment. If drafts are present, position the equipment

to avoid them.

• Allow the instruments to warm up for at least 30 minutes before performing the test.

Note that you must wait at least 15 minutes after wiring the thermocouple.

• Always use terminal covers.
• If the measured values are questionable, record 50 to 100 data, then make a

determination based on the average or median value.

• The MX110-UNV-H04 modules have a separate RJC for each channel, so the criteria

must be met on each channel (1 through 4).

• Measurement fluctuation errors must be prevented as much as possible by adopting

the following setting conditions.
Measurement interval: 100 ms; Filter: 0

• Confirm that the fluctuation among 50 data meets the criterion.

Criterion: 0.2°C or less

If the test results do not meet the criteria, reconfirm the items above.

Procedure

Perform measurement using the Type-T thermocouple at 0°C.
Confirm that the measured values meet the criterion.
Criterion: 0±0.4°C or less

Causes of Failure

If the test results do not meet the criteria, the following causes may apply.
• Faulty RJC board connection
• Defective RJC board (B8720TH)