2 tests of the main module (mx100), Tests of the main module (mx100) -2, Tests – Yokogawa PC-Based MX100 User Manual

Page 31: Instruments used, Before testing, Procedure

Advertising
background image

2-2

SM MX100-01E

2.2 Tests of the Main Module (MX100)

Tests

2.2.1 Protective Earth Continuity Test
2.2.2 Insulation Resistance Test
2.2.3 Withstanding Voltage Test
2.2.4 Voltage Frequency Detection Function
2.2.5 Battery Backup Function
2.2.6 Module Communication Function
2.2.7 LED Display Test
2.2.8 CF Card Test

Instruments Used

Instrument

Description

Insulation resistance tester

YOKOGAWA 2407 or equivalent
A device that can measure 100 MΩ or more at 500 VDC

Withstanding voltage test

Kikusui TOS-9201 or equivalent
A device that can measure 1.5 kVAC for 1 minute with a limit

current of 10 mA

Earth continuity tester

Kikusui TOS-6200 or equivalent

AC stabilized power supply

A device capable of outputting AC100 V, and switching between 50

and 60 Hz

Input module or base plate (for test)

Before Testing

Initialize the memory before beginning any test. To initialize the memory, use the API for
MX100/DARWIN (the MX190 application software), or perform the following procedure
using the dip switches.

Procedure

1.

Turn OFF the power to the MX100.

2.

Set the dip switches as shown in the figure below. (Turn switch 5 OFF.)

1

2

3

4

5

6

7

8

ON

3.

Turn ON the power to the MX100.

4.

Check that bF is displayed on the 7-segment LED, then turn OFF the power to the

MX100 again.

5.

Set the dip switches as shown in the figure below (for normal operation, all

switches are turned ON.)

1

2

3

4

5

6

7

8

ON

Advertising