6 between-channel error test, 7 di contact input test, Procedure – Yokogawa PC-Based MX100 User Manual

Page 53

Advertising
background image

2-24

SM MX100-01E

2.6.6

Between-Channel Error Test

Procedure

1.

Connect the I, A, B, and C terminals on all channels to a voltage generator or

variable resistor, then confirm that the measured values on each channel meet
the criteria.

Set the measurement interval on the module under test to 500 msec.

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15

/I

+/A
-/B

/C

/I

+/A
-/B

/C

/I

+/A
-/B

/C

4CH

6CH

5CH

1CH

2CH

3CH

DC Voltage

generator

or variable

resistor

2.

Confirm that the between-channel error vales are less than the criteria given

below.

Range

Input

Criteria

20 mV

Short input H to L

0.000 mV ±0.005 mV

20 Ω

20 Ω

20.000 Ω ±0.017 Ω

2.6.7

DI Contact Input Test

Connect the resistances below between the A and B terminals on any two channels, then
confirm that the criteria are met.
Set the measurement interval on the module under test to 500 msec.

Range

Load

Criteria

DI CONTACT

2 kΩ

1

DI CONTACT

Connect 200 kΩ + 0.022 μF in parallel 0 or 20 kΩ

2.6 Tests of the 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module (MX110-V4R-M06)