1 insulation resistance test, 2 withstanding voltage test – Yokogawa PC-Based MX100 User Manual

Page 73

Advertising
background image

2-44

SM MX100-01E

2.10.1 Insulation Resistance Test

Install the module on the MX100, MW100, or equivalent before performing this test.
Perform the withstanding voltage test before this test.
Wire the devices as in the figure below, then confirm that the criteria are met.

Inpout terminal to earth terminal on

MX100, MW100, or equivalent

To the earth terminal on the MX100,

MW100, or equivalent device

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15

Insulation resistance

tester or

Withstanding voltage

tester

Terminals Measured

Criteria

Input terminal to earth terminal on the MX100, MW100, or

equivalent

20 MΩ or more (at DCV 500 V)

2.10.2 Withstanding Voltage Test

Install the module on the MX100, MW100, or equivalent before performing this test.

Note

▪ In order to ensure continuity from the MX100, MW100, or equivalent’s earth terminal

to the module under test’s heat sink, you must fasten the module with screws after
installing it onto the base plate.

▪ When applying voltages, ensure that the rising and falling times are both at least five

seconds.

Wire the instrument in the same manner as the insulation resistance test, then confirm
that the results satisfy the criteria below.

Terminals Measured

Criteria

Input terminal to earth terminal on the MX100, MW100, or

equivalent

2300 Vrms AC for 1 minute
Leakage current of 2 mA or less