Procedure – Yokogawa PC-Based MX100 User Manual

Page 46

Advertising
background image

2-17

SM MX100-01E

Testing

1

2

3

4

5

6

7

2.5.5

Measurement Accuracy Test

Procedure

1.

Turn the power ON then OFF.

2.

Turn the power ON again, then allow the instrument to warm up for at least 10

minutes.

3.

Adjust so that the wires connected to the A, B, and b terminals have the same

resistance value.

Set channel 2 as the measurement channel, and the measurement interval and
duration as follows:

Measurement Interval

Duration

100 ms

5 seconds or more

500 ms

10 seconds or more

4.

Confirm that the measured results are within the judgment criteria range in the

chart below.

100 ms Interval

Range

Input

Specifications

2 V

+2.0000 V
0.0000 V
-2.0000 V

±(0.1% of rdg. + 10 digits)

500 ms Interval

Range

Input

Specifications

20 mV

+2.0000 V
0.000 mV
-20.000 mV

±(0.05% of rdg. + 5 digits)

60 mV

+60.00 mV
0.00 mV
-60.00 V

±(0.05% of rdg + 2 digits)

200 mV

+200.00 mV
0.00 mV
-200.00 mV

±(0.05% of rdg + 2 digits)

2 V

+2.0000 V
0.0000 V
-2.0000 V

±(0.05% of rdg + 5 digits)

6 V

+6.000 V
0.000 V
-6.000 V

±(0.05% of rdg + 2 digits)

20 V

+20.000 V
0.000 V
-20.000 V

±(0.05% of rdg + 2 digits)

100 V

+100.00 V
0.00 V
-100.00 V

±(0.05% of rdg + 2 digits)

Pt100

18.52 Ω/-200.0°C
100.00 Ω/0.0°C
313.71 Ω/+600.0°C

±(0.05% of rdg + 0.3°C)

Cu10 GE

9.036 Ω/-0.0°C

±(0.1% of rdg + 2°C)

Pt100 High Resolution 100.00 Ω/0.00°C

157.33 Ω/150.00°C

±(0.05% of rdg + 0.3°C)

Cu10 GE

1.326 Ω/-200.0°C

±(0.1% of rdg + 2°C)

High Resolution

9.036 Ω/0.0°C
20.601 Ω/+300.0°C

2.5 Tests of the 10-CH, Medium-Speed Universal Input Module (MX110-UNV-H10)