2 insulation resistance test, 3 withstanding voltage test – Yokogawa PC-Based MX100 User Manual

Page 32

Advertising
background image

2-3

SM MX100-01E

Testing

1

2

3

4

5

6

7

2.2.1

Protective Earth Continuity Test

Procedure

1.

Wire the instruments as shown below.

Earth continuity

tester

Protective earth

terminal

MX100

(main unit)

Functional

earth terminal

I–

V–

I+

V+

Protective earth terminal

L

N

G

When using an inlet

When using a screw terminal

Center pin

Right-most pin

2.

Measure the resistance between the protective earth terminal and functional earth

terminal. Use 4 wires for this measurement.

Items to Be Checked

That the resistance value does not exceed 0.1 Ω when tested at 60 Hz, 30 A for 10
seconds.

2.2.2

Insulation Resistance Test

Procedure

Perform a test using a DC 500 V insulation resistance tester and confirm that the
following criteria are met.

Terminals Measured

Power Switch

Criteria

AC power terminal to functional earth termina

l*

(Short between L and N)

ON

20 MΩ or higher

Ethernet input terminal to functional earth terminal
(short all pins)

OFF

20 MΩ or higher

* You may use the functional earth terminal as a protective earth terminal.

2.2.3

Withstanding Voltage Test

Procedure

Perform the withstanding voltage test, and confirm that the following criteria are met and
that no abnormalities occur.

Terminals Measured

Power Switch

Criteria

AC power terminal to functional earth terminal

*

(Short between L and N)

ON

1.5 kVAC for 1 minute
Leakage current of 5

mA or less

* You may use the functional earth terminal as a protective earth terminal.