1 protective ground continuity test, 2 insulating resistance test, 3 withstanding voltage test – Yokogawa PC-Based MX100 User Manual

Page 35: Procedure

Advertising
background image

2-6

SM MX100-01E

2.3.1

Protective Ground Continuity Test

Procedure

1.

Wire the instruments as shown below.

I–

V–

I+

V+

Earth continuity

tester

Protective earth

terminal

MW100

(main unit)

Functional

earth terminal

Protective ground terminal

L

N

G

When using an inlet

When using a screw terminal

Center pin

Right-most pin

2.

Measure the resistance between the protective ground terminal and functional

ground terminal. Use 4 wires for this measurement.

Items to Be Checked

That the resistance value does not exceed 0.1 Ω when tested at 60 Hz, 30 A for
10 seconds.

2.3.2

Insulating Resistance Test

Procedure

Perform a test using a DC 500 V insulation resistance tester and confirm that the
following criteria are met.

Terminals Measured

Power Switch

Criteria

Power terminal to functional ground terminal*

ON

20 MΩ or higher

MW100 AC power supply specifications

(short between L and N)
MW100 DC power supply specifications

(short +)

* You may use the functional ground terminal as a protective ground terminal.

2.3.3

Withstanding Voltage Test

Procedure

Perform the withstanding voltage test, and confirm that the following criteria are met and
that no abnormalities occur.

Terminals Measured

Power Switch

Criteria

AC power supply terminal to functional ground

terminal*

ON

1.5 kVAC for 1

minute

MW100 AC power supply specifications

(short between L and N)

Leakage current of 5

mA or less

MW100 DC power supply specifications

(short +)

* You may use the functional ground terminal as a protective ground terminal.