Module (MX115-D05-H10)

Tests

2.8.1 Insulation Resistance Test
2.8.2 Withstanding Voltage Test
2.8.3 Digital Input Test
2.8.4 ID Check

Instruments Used

Instrument

Standards, Recommended Products Etc.

MX100, MW100, or equivalent

MX100 or MW100 Main Module
MX150 Base Plate

PC

Ensure that the software required for the test is installed

DC voltage generator

Accuracy: ±1% + 10 mV or less

Variable resistor or fixed resistance

Accuracy: ±1% or less

Insulation resistance tester

DC500 V

Withstanding voltage tester

AC 1 kV–3 kV

Front terminal

FRONT TERMINAL ASSY (B8720BF)

*

Screw terminal

10-CH screw terminal block (772061)

Cable

Combination cable (772062-050 or 772062-100)

* Regarding the front terminal
If /NC is included in the model name of the assembly, use B8720BF.
If /NC is not included in the model name of the assembly, do not use B8720BF.
† Regarding the screw terminal block and cable
Used for confirmation of the ID setting. If the ID matches, you can substitute with a device

whose connector has been altered.

‡ That the module settings and measured values can be read in.

Cautions When Performing This Test

• Connect to the PC only for the digital input test.
• If the /NC option is installed, use B8720BF during the test.
When the test is complete, be sure you remove the B8720BF.
If option /NC is not installed, do not use the B8720BF.

2.8.1

Insulation Resistance Test

For this test, the Ten-Channel, High-Speed Digital Input Module is installed with the
MX100, MW100, or equivalent device.

Procedure

1.

The insulation resistance between the terminals indicated below are tested. When

performing the test, you must fasten the module being tested to the MX150 base
plate jig with screws.

Input terminals: short the input terminals on all channels

Earth terminals: the unit or equivalent’s functional earth terminal

2.

Confirm that the criterion below is met.

Input terminal to earth terminal: an insulation resistance of 20 MΩ or more at

DC500 V