4 id check, 5 bridge voltage accuracy test, Wiring diagram, h to l – Yokogawa PC-Based MX100 User Manual

Page 58: Testing, Measured waveform, 7 tests of the 4-ch strain input module

Advertising
background image

2-29

SM MX100-01E

Testing

1

2

3

4

5

6

7

Wiring diagram, H to L

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15
16

ON

1

2

3
4

5

ON

1

2

3
4

5

ON

1

2

3
4

5

ON

1

2

3
4

5

A/+V
B/ L
C/-V
D/ H

CH1

A/+V
B/ L
C/-V
D/ H

CH2

A/+V
B/ L
C/-V
D/ H

CH3

A/+V
B/ L
C/-V
D/ H

CH4

DC voltage

generator

CH1

CH2

CH3

CH4

D

H

A1002JC

A1002JC

A1002JC

A1002JC

Pin name
Signal name

Pin name
Signal name

B

L

DC voltage

generator

2.7.4

ID Check

This test confirms whether the module is correctly identified when module information is
obtained from the MX100, MW100, or equivalent.
(If the OFF/ON cycle is not performed on the MX100, MW100, or equivalent, the ID
information is not updated. )

2.7.5

Bridge Voltage Accuracy Test

Wire as indicated below, then measure the bridge voltage on the waveform device.
Confirm that the bridge voltage satisfies the criteria.

Measured CH

Any

Load resistance criterion

100 Ω ±1%, 0.1W or more

Bridge voltage criterion

2 V ±0.05 V

Measured waveform

Bridge voltage

0V

2.7 Tests of the 4-CH Strain Input Module

Advertising
This manual is related to the following products: