2-10

SM MX100-01E

2.4.2

Withstanding Voltage Test

For this test, the Four-Channel, High-Speed Universal Input Module is installed with the
MX100, MW100, or equivalent device. Perform this test after the test described in section
2.4.1, “Insulation Resistance Test.”

Procedure

Perform the withstanding voltage test on the same locations indicated for the

insulation resistance test (input terminal to earth, and between input terminals), then
confirm that the results meet the criteria below.

Be sure to attach the module and base plate when performing this test.

Measured Item

Criteria

Input terminal to earth terminal

AC3700 Vrms for 1 minute, with leakage of 2 mA or less

Between input terminals

AC2300 Vrms for 1 minute, with leakage of 1 mA or less

2.4.3

Excessive Input Test

Perform this test while in measurement mode.
Set the measurement interval to an arbitrary value greater than 50 ms.

Procedure

Set the range for all channels to DC20 mV.

1.

Connect all A (1) and B (2) terminals separately on channels 1 through 4 as

shown in the figure below.

A 1

B 2

b 3

3ch

A 1

B 2

b 3

1ch

A 1

B 2

b 3

4ch

A 1

B 2

b 3

2ch

DC voltage

generator

2.

Apply the voltages below between the A and B terminals, and confirm that no

abnormalities occur.

Note

In the same manner as described in “Setting All Channels to Pt100/1 mA High Resolution

Range” below, connect terminals B and b (2, 3) then perform the excessive input test.

Applied Voltage

Duration

+20 V

10 seconds

-20 V

10 seconds

2.4 Tests of the 4-CH, High-Speed Universal Input Module (MX110-UNV-H04)