8 initial balance error test – Yokogawa PC-Based MX100 User Manual

Page 63

Advertising
background image

2-34

SM MX100-01E

2.7.8

Initial Balance Error Test

1.

Connect the Cal Box to the measurement channels.

2.

Enter the following settings.

Measured CH

Any (CH2 recommended)

Measurement interval

200 ms

Filters

OFF

Duration

10 s or more

Dip switch setting (for -B12 and -B35) 4-gauge method

ON

1 2 3 4 5

SW1 to SW4: OFF
SW5: ON

3.

Perform initial balancing using the indicated initial balance values for each

measuring range.

4.

Apply the following input values and confirm that the measured values meet the

criteria.

Test Range

Test Points

Initial balance

Specification

Measured value Setting Output value Setting

2000 µSTR

+2000 µSTR

No. 2

+10000 µSTR No.12

±0.5% of Range

0 µSTR

No.12

2000 µSTR

No. 3

-10000 µSTR No.13

0 µSTR

No.13

20000 µSTR

+20000 µSTR

No. 4

+10000 µSTR No.12

±0.3% of Range

0 µSTR

No.12

20000 µSTR

No. 5

-10000 µSTR No.13

0 µSTR

No.13

For instructions on setting test points, see “Cal Box Output Settings.”