5 pulse interval accuracy test, Mw100 settings – Yokogawa PC-Based MX100 User Manual

Page 75

Advertising
background image

2-46

SM MX100-01E

<MW100 Settings>

Referring to the MW100 user’s manual for browser operations, or by using
communication commands, power cycle the MW100 main unit or equivalent, then
measure the output on all channels.

Output Range Settings

Item

Setting

Remarks

Output type

PWM

Output method

COM

Command output

Output range

1 ms

Span

Need not specify

Pulse interval

10

Preset value

0.000%

Duty 0%

Reference channel

Not required

Output Operation Setting

Item

Setting

Power ON

PRESET

Error occurence

PRESET

Criteria (for the MX100 and MW100)

PWM Output Setting

Criteria

100.0 %

3.8 V or more

2.10.5 Pulse Interval Accuracy Test

Wire the units as in the figure below, enter all settings (PWM output settings), then
confirm that the criteria are met.
The figure below shows an example of wiring CH5. Actual measurment should be
performed on all channels.
Use an external power supply of 5 V, load resistance of 100 Ω, and a counter trigger level
of 2.5 V.

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15

+
-

100Ω

Counter +

-

5V

V+

-

5Ch

V+

-

6Ch

V+

-

7Ch

V+

-

8Ch

+
-

Vext

+
-

Vext

V+

-

4Ch

V+

-

3Ch

V+

-

2Ch

V+

-

1Ch

Puls interval check and puls duty check wiring diagram

DC

Power

supply

2.10 8-CH, Medium-Speed PWM Output Module