5 output accuracy test – Yokogawa PC-Based MX100 User Manual

Page 69

Advertising
background image

2-40

SM MX100-01E

<MX100>

1.

Enter settings in the table below to specify output at power ON using the MX100

standard software or the MX100/DARWIN API. Power cycle the MX100 main
unit or equivalent, then measure the output on all channels. (Without PC
communications.)

Enter the same settings on all channels.
When using the API for settings, enter settings that are equivalent to the following.

Item

Setting

Description/Notes

Enable/disable

Enabled

Action

User output

Voltage/current

Voltage

Span

Need not specify -10–10 V

Output upon power ON

Preset value

Sets the power ON operation as the preset.

Output upon error

Preset value

Sets the output upon error as the preset.

Preset value

+11 V

Sets +11 V as the preset value.

Other

Not required

2

Confirm that the output voltages of all channels are within the allowable ranges

shown below.

Output Terminal

Allowable Range

Voltage

11 V ±32 mV

Current

22 mA–24.6 mA

<MW100>

1.

Referring to the MW100 user’s manual for browser operation, or by using

communication commands, set the output at power ON according to the table
below. Power cycle the MW100 main unit or equivalent, then measure the output
on all channels.

Output Range Settings

Item

Setting

Description/Notes

Output Type

AO

Output method

COM

Command output

Output range

10 V

Span

Need not specify -10–10 V

Preset value

+11 V

Reference channel

Need not specify

Output Operation Setting

Item

Setting

Power ON

PRESET

Error occurence

PRESET

2

Confirm that the output voltages of all channels are within the allowable ranges

shown below.

Output terminal

Allowable Range

Voltage

11 V ±32 mV

Current

22 mA – 24.6 mA

2.9.5

Output Accuracy Test

Measurement Conditions

After calibration, if you do not turn OFF the power at least once, even if you perform
measurement correctly the calibration data may not be recorded accurately. Therefore,
power cycle the unit before performing the test.
Perform measurement on all channels.

2.9 Tests of the 8-CH Analog Output Module

Advertising