2.12.5 Reference Junction Compensation Accuracy Test

Always use terminal covers.

Procedure

1.

Allow a warm-up of 30 minutes or more.

2.

Measure the 0°C standardized temperature tank at the range for Type-T

thermocouples.

3.

Measure at all channels, and confirm that the result lies within 0°C ± 0.5°C.

2.12 Tests of the 10-CH Screw Terminal Block (772061)