2.6.1

Insulation Resistance Test

Attach the MX110 6-CH Medium-Speed Four-Wire RTD Resistance Input Module to the
MX100, MW100, or equivalent before performing this test.

Procedure

Wire the instrument as shown in the figure below, measure the insulation resistance, then
confirm that the results meet the criteria below.

Measured Item

Criteria

Input terminal to earth terminal on the MX100, MW100, or

equivalent

20 MΩ or more (at DCV 500 V)

Between input terminals

20 MΩ or more (at DCV 500 V)

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15

/I

+/A
-/B

/C

/I

+/A
-/B

/C

/I

+/A
-/B

/C

4CH

6CH

5CH

1CH

2CH

3CH

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15

/I

+/A
-/B

/C

/I

+/A
-/B

/C

/I

+/A
-/B

/C

4CH

6CH

5CH

1CH

2CH

3CH

Insulation resistance

tester or

Withstanding voltage

tester

Insulation resistance

tester or

Withstanding voltage

tester

Input terminal to earth terminal on the jig

Between input terminals

2.6.2

Withstanding Voltage Test

Attach the 6-CH Medium-Speed Four-Wire RTD Resistance Input Module to the MX100,
MW100, or equivalent before performing this test.

Procedure

Wire the instrument in the same manner as in section 2.6.2, “Insulation Resistance Test,”
then confirm that the results satisfy the criteria below.
For the withstanding voltage test, in order to ensure continuity from the MX100, MW100,
or equivalent’s earth to the module’s heat sink, you must fasten the module with screws
after installing it onto the base plate.

Measured Item

Criteria

Input terminal to earth terminal on the

MX100, MW100, or equivalent

Current leakage of 2 mA or less at 3700 Vrms AC for 1

minute

Between input terminals

Current leakage of 1 mA or less at 1000 Vrms AC for 1

minute

When applying voltages, ensure that the rising and falling times are both at least five
seconds.